8.4 Manual control - Functions

■■■■■■■■■■■■■■■■■■■■■■

1616

■■■■■■■■

tiamo 2.4

Gas flow

Inlet

Selection of the carrier gas inlet.

Selection

Pump | Valve

Default value

Pump

Pump
Use of ambient air as carrier gas.
Valve
Use of gas from a compressed gas cylinder.

Flow rate

874
Input range

10 - 150 mL/min

Default value

50 mL/min

Gas type

Gas type of the carrier gas.

Selection

Air | Nitrogen | Other gas

Default value

Air
In the case of the 874 USB Oven Sample Processor,
a gas flow factor has to be specified for the carrier
gas used if Other gas is selected. The gas flow fac-
tor is used for the correct measurement of the flow
rate.

Gas flow factor

Factor for the correct determination of the gas flow. This parameter is dis-
played only if Gas type = Other gas. For air and nitrogen, the appropri-
ate gas flow factor is inserted automatically. The gas flow factor values for
other commonly used carrier gases can be found in the table Gas flow
factors of various gases
.

Input range

0.001 - 2.000 (Increment: 0.001)

Default value

1.000

Table 45

Gas flow factors of various gases

Gas type

Gas flow factor

Argon

0.950

Helium

0.300

Air

1.000