Pause 2

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.

Input range

0 - 999999 s

Default value

0 s

5.6.3.4.5.4

SET Ipol - Control parameters

Tab: Method

SET Ipol Properties... Control parameters

Command name

Name of the command.

Entry

25 characters

The control parameters can be set for each endpoint separately.

Endpoint 1

EP1 with

Measured value for the first endpoint.

Titrando, 855
Input range

-2000.0 - 2000.0 mV

Selection

off

Default value

off

Titrino
Input range

-2000 - 2000

Selection

off

Default value

off

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.

Selection

slow | optimal | fast | User

Default value

optimal