5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1180
■■■■■■■■
tiamo 2.4
Titrando, 855
Input range
0.00000 - 9999.99 mL
Default value
100.000 mL
Selection
off
off
Means that no stop will take place.
Titrino
Input range
0 - 9999.99 mL
Default value
100.00 mL
Selection
off
off
Means that no stop will take place.
Stop time
Stops when the preset time has elapsed since the start of the dosing.
Input range
0 - 999999 s
Selection
off
Default value
off
off
Means that no stop will take place.
Filling rate
Rate at which the dosing cylinder is to be filled after dosing. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing device used. If the entered filling rate is too high for the selected
dosing device, it will automatically be reduced during dosing to the largest
possible value.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Default value
maximum mL/min
Titrino
Input range
0.01 - 150.00 mL/min
Default value
maximum mL/min
NOTE
The filling rate should be reduced for viscous liquids.