6 stat u - stop conditions, Stop conditions – Metrohm tiamo 2.4 Manual User Manual
Page 921
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
905
Titrando, 855
Input range
0.01 - 166.00 mL/min
Selection
off
Default value
off
Titrino
Input range
0.01 - 150.00 mL/min
Selection
off
Default value
off
5.6.3.8.7.6
STAT U - Stop conditions
Tab: Method
▶ STAT U ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters
Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.
Stop volume
Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Titrando, 855
Input range
0.00000 - 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Input range
0.000 - 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Stop when the entered time has elapsed after the start conditions have
been fulfilled.
Input range
0 - 999,999 s
Selection
off
Default value
off