6 stat u - stop conditions, Stop conditions – Metrohm tiamo 2.4 Manual User Manual

Page 921

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

905

Titrando, 855
Input range

0.01 - 166.00 mL/min

Selection

off

Default value

off

Titrino
Input range

0.01 - 150.00 mL/min

Selection

off

Default value

off

5.6.3.8.7.6

STAT U - Stop conditions

Tab: Method

STAT U Properties... Stop conditions

Command name

Name of the command.

Entry

25 characters

Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.

Stop volume

Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.

Titrando, 855
Input range

0.00000 - 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Input range

0.000 - 9999.99 mL

Default value

100.00 mL

Selection

off

Stop time

Stop when the entered time has elapsed after the start conditions have
been fulfilled.

Input range

0 - 999,999 s

Selection

off

Default value

off

Advertising