6 Configuration

tiamo 2.4

■■■■■■■■

1381

Statistics
Statistics data will only be shown if the titer has been assigned by the titer
method as a mean value of a multiple determination. If the titer has been
determined with a single determination or entered manually then no sta-
tistics values will be shown.

Number of titer determinations (n)

Number of determinations used to determine the titer.

Abs. standard dev. (s abs)

Absolute standard deviation for the titer determination in the unit of the
titer.

Rel. standard dev. (s rel)

Relative standard deviation for the titer determination in %.

Monitoring of titer validity
on | off (Default value: off)

If this option is activated, then the validity of the titer will be monitored.

Validity of titer

Validity of the titer in days. If a value is entered here, then the date for the
Next titer determination will be automatically adapted.

Input range

0 - 999 days

Default value

999 days

Next titer determination

Date on which the next titer determination is to be carried out. The date

can be selected by clicking on

in the Select date dialog window (see

Chapter 2.5.1, page 90). After the date has been entered, the Validity
field will be adapted automatically.

Default value

   Date titer det.+ 999 days

Message
The message options can only be edited if monitoring is switched on.

Message by e-mail

on | off (Default value: off)

If this option is enabled, the message is sent to the address defined under
[E-mail]. The message is sent in text format.