5 Method

tiamo 2.4

■■■■■■■■

833

Selection

without diaphragm | with diaphragm

Default value

without diaphragm

Generator current

Polarization current at the generator electrode.

for generator type = without diaphragm
Selection

100 mA | 200 mA | 400 mA | auto

Default value

400 mA

auto
Means that the current is automatically adapted to the conductivity of
the reagent and that near the endpoint the current will be controlled
at smaller values.

for generator type = with diaphragm
Selection

100 mA | 200 mA | 400 mA | auto

Default value

auto

auto
Means that the current is automatically adapted to the conductivity of
the reagent and that near the endpoint the current will be controlled
at smaller values.

NOTE

For generator electrodes of the type with diaphragm, it is recommen-
ded to set the Generator current to auto.

For generator electrodes without diaphragm it is recommended to
set the Generator current to 400 mA.

Stirrer

Stirrer

Selection of the stirrer.

851, 852
Selection

1 | 2 | 3 | 4 | off

Default value

1

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.