5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

718

■■■■■■■■

tiamo 2.4

Parameters
The parameters for the SET U command are set on the following eight
tabs:

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

Control parameters
Control parameters for endpoint 1 and endpoint 2 (measured values
for the endpoints, titration rate, user-defined parameters, stop crite-
rion).

Titration parameters
Parameters for the run of the titration.

Stop conditions
Definition of conditions which cause the titration to stop.

Conditioning
Enable/disable conditioning. Definition of start drift, drift correction
and stop conditions.

Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.

Additional measured values
Definition of additional measured values of other measuring com-
mands, which are saved as additional columns in the measuring point
list.

Command variables
The following command variables are generated by the SET U command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':

Identification

Description

.BLV

Blank value of the sensor used for the command (only
for ISE sensors)

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started