5 Method

tiamo 2.4

■■■■■■■■

1021

Selection

712 Conductometer | 856 Conductivity Mod-
ule

Default value

856 Conductivity Module

Sensor

Sensor

Selection of a sensor of the Conductivity sensor type from the sensors
configured in the sensor table.

Selection

Sensor name | Conductivity sensor

Default value

Conductivity sensor

Temperature measurement

Type of temperature measurement:

856
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
under Temperature on the Measuring parameters tab will be
used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.

Stirrer

NOTE

Is displayed only if the 856 Conductivity Module has been selected as
device type.

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1