5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

864

■■■■■■■■

tiamo 2.4

However, in order to ensure uninterrupted dosing, the following points
must be taken into account:

Keeping filling times short
Use the highest possible filling rate in order to keep the filling times as
short as possible. Take the viscosity and density of the liquid into
account.

Filling rate with different cylinders
When you use two dosing devices with different cylinder volumes (dos-
ing cylinder 2 > dosing cylinder 1), then the minimum filling rate of the
larger cylinder 2 must be:
Filling rate 2

≧ dosing rate 1 • (V

dosing cylinder 2

/ V

dosing cylinder 1

)

Rule for dosing rate
The dosing rate must not exceed 75% of the filling rate of the smaller
cylinder. This corresponds to the following values (at maximum filling
rate):

Cylin-

der vol-

ume

maximum dosing rate

max. flow rate

Exchange

unit

Dosing

unit

Exchange

unit

Dosing

unit

1 mL

2.25 mL/min

---

approx. 130

mL/h

---

2 mL

---

5.00

mL/min

---

approx. 300

mL/h

5 mL

11.25

mL/min

12.50

mL/min

approx. 670

mL/h

approx. 750

mL/h

10 mL

22.50

mL/min

25.00

mL/min

approx. 1.3 L/h

approx. 1.5

L/h

20 mL

45.00

mL/min

50.00

mL/min

approx. 2.7 L/h

approx. 3.0

L/h

50 mL

112.50

mL/min

124.50

mL/min

approx. 6.7 L/h

approx. 7.5

L/h

5.6.3.8.6

STAT pH

5.6.3.8.6.1

STAT pH - Overview

Dialog window: Method

STAT pH Properties... STAT pH - 'Command name'

Command for STAT titrations with measured value pH.