5 Method
tiamo 2.4
■■■■■■■■
629
Input range
0 - 999999 s
Default value
0 s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Input range
0 - 999999 s
Default value
1 s
Start volume
Start volume
Volume to be added before the start of the titration at the dosing rate
indicated.
Titrando, 855
Input range
0.00000 - 9999.99 mL
Default value
0.00000 mL
Titrino
Input range
0.00 - 999.99 mL
Default value
0.00 mL
Dosing rate
Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Input range
0.01 - 150.00 mL/min
Selection
maximum
Default value
maximum
Start measured value
NOTE
Only displayed for Titrando and 855.