5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

926

■■■■■■■■

tiamo 2.4

The following additional methods for evaluation of titration curves can be
activated and defined on this tab:

Fixed endpoint evaluation

Fixed endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other mea-
sured quantities will be interpolated with a fixed measured quantity (Mea-
sured value
or Volume) for the fixed endpoint from the measuring point
list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.9.10, page 929).

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.9.10, page 929).

[Delete]

Delete the selected line.

Minimum evaluation
on | off (Default value: off)

If this option is activated, then the associated volume for the minimum
measured value will be interpolated from the measuring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Only editable for Minimum evaluation = on, otherwise disabled.

Input range

0.1 - 20.0 °C/mL

Default value

1.0 °C/mL