5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1196

■■■■■■■■

tiamo 2.4

Titrando, 855
Input range

0.01 - 166.00 mL/min

Default value

maximum mL/min

Titrino
Input range

0.01 - 150.00 mL/min

Default value

maximum mL/min

NOTE

The filling rate should be reduced for viscous liquids.

5.6.6.3.5.5

DOS U - Monitoring

Tab: Methods

DOS U Properties... Monitoring

Command name

Name of the command.

Entry

25 characters

The following monitoring processes can be switched on and defined on
this tab:

Monitoring Measured value

Monitoring Measured value

on | off (Default value: off)

If this option is switched on then the measured value will be monitored
and any exceeded limits will be entered in the measuring point list.

Lower limit

Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is trig-
gered.

Titrando, 855
Input range

-2000.0 - 2000.0 mV

Default value

-2000.0 mV

Titrino
Input range

-2000 - 2000 mV

Default value

-2000 mV