Metrohm tiamo 2.4 Manual User Manual

Page 605

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

589

Window = Measured value
Input range

-20.000 - 20.000 pH

Default value

-20.000 pH

Window = Volume
Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Window = Time
Input range

0 - 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Window = Measured value pH
Input range

-20.0 - 20.0 pH

Default value

20.0 pH

Window = Volume
Input range

0.00000 - 9999.99 mL

Default value

9999.99 mL

Window = Time
Input range

0 - 999999 s

Default value

999999 s

Gran evaluation

NOTE

This evaluation method is possible only with Titrando and 855.

Gran evaluation

on | off (Default value: off)

If this option is enabled, then titration curves will be evaluated in accor-
dance with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 932).

Procedure

Selection of the Gran procedure.

Selection

Normalized | Standard

Default value

Normalized