Metrohm tiamo 2.4 Manual User Manual

Page 66

Advertising
background image

2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

tiamo 2.4

Identifica-
tion

Description

Commands

.SAN

Current absolute swing angle of
the robotic arm in ° (entry upon
exiting the command)

MOVE, SWING

.SLO

Electrode slope of the sensor used
for the command (in % for pH
sensor or mV for ISE sensor) or
electrode slope calculated from
the calibration (for STDADD and
CAL LOOP)

DET pH, DET U,
MET pH, MET U,
SET pH, SET U,
STAT, MEAS pH,
MEAS U, MEAS
T, MEAS Conc,
STDADD, CAL
LOOP, DOS, ELT
LOOP

.SME

Start measured value (measured
value after processing the start
conditions) in the unit of the
measured value

DET, MET, SET,
KFT, KFC, BRC,
STAT

.SPO

Current external position (entry
when ending the command); 0
means invalid position

SWING

.STC {x}

Concentration of the standard x
(1 - 50) used for the command

CAL LOOP Opt

.STE

Start temperature (temperature
after processing the start condi-
tions) in °C

DET, MET, SET,
KFT, KFC, BRC,
STAT

.STY

Type of stop with which the com-
mand was stopped: 1 = normal;
0 = manual or after error

DET, MET, SET,
KFT, STAT,
MEAS, CAL
Cond, CAL
MEAS, CAL Spec,
STDADD, DOS,
ELT MEAS

.SVA

Start volume absolute (volume
that was added according to the
start condition "start volume") in
mL

DET, MET, SET,
KFT, STAT

.SVM

Start volume measured value (vol-
ume that was added according to
the start condition "start mea-
sured value") in mL

DET, MET

Advertising