5 Method

tiamo 2.4

■■■■■■■■

793

Input range

0 - 999999 s

Default value

0 s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.

Input range

0 - 999999 s

Default value

1 s

Pause 1

Pause 1

Waiting time, e.g. for the electrode to settle down, before a start volume
is added.

Input range

0 - 999,999 s

Default value

0 s

Start volume

Start volume

Volume to be dosed with the indicated dosing rate before the start of the
titration.

Titrando
Input range

0.00000 - 9,999.99 mL

Default value

0.00000 mL

Titrino
Input range

0.00 - 999.99 mL

Default value

0.00 mL

Dosing rate

Rate at which the start volume is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.

Titrando
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum