5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1186

■■■■■■■■

tiamo 2.4

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 934)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.6.3.5

DOS U

5.6.6.3.5.1

DOS U - Overview

Dialog window: Method

DOS U Properties... DOS U - 'Command name'

Command for controlled dosing with measured value U.

Devices
This command can be executed with the following devices:

Titrando: 835, 836, 842, 857, 902, 906, 907

Titrino: 718, 736, 751, 799

pH/Ion Meter: 867

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Parameters
The parameters for the command DOS U are configured in the following
5 tabs:

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

Dosing parameters
Parameters for the dosing settings.

Stop conditions
Definition of conditions which cause the stop of the dosing.

Monitoring
Definition of measured value and temperature monitoring.

Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional columns in the measuring point
list.