Metrohm tiamo 2.4 Manual User Manual

Page 833

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

817

has been reached. The larger the control range, the slower the titration.
Outside the control range, dosing is carried out continuously at Max.
rate
.

Titrando
Input range

0.01 - 200.0 µA

Default value

40.0 (slow), 10.0 (optimal), 5.0 (fast) µA

Selection

off

off
The entire measured value range is defined as control range. The mini-
mum titration rate is applied throughout the entire titration.

Titrino
Input range

0.1 - 200.0 µA

Default value

40.0 (slow), 10.0 (optimal), 5.0 (fast) µA

Selection

off

off
The entire measured value range is defined as control range. The mini-
mum titration rate is applied throughout the entire titration.

Max. rate

Rate at which dosing is carried out outside of the Control range. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.

Titrando
Input range

0.01 - 166.00 mL/min

Default value

1.00 (slow), maximum (optimal), maximum
(fast) mL/min

Selection

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Default value

1.00 (slow), maximum (optimal), maximum
(fast) mL/min

Selection

maximum

Min. volume increment

This parameter determines the addition rate with which dosing is carried
out at the very beginning and at the end of the titration. The smaller the
increment that is selected, the slower the titration and the higher the
accuracy.

Input range

0.1 - 9.9 µL

Default value

minimum (slow), minimum (optimal), 5.0
(fast) µL