5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

530

■■■■■■■■

tiamo 2.4

descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.

Evaluation with measured value window (Ipol)

EP criterion

Criterion for the recognition of equivalence points which is valid for all
windows. Equivalence points whose ERC is smaller than the set EP crite-
rion will not be recognized.

Titrino
Input range

0 - 200

Default value

5

[New]

Opens the Measured value window # dialog window, in which the
parameters for a new window can be entered.

[Properties]

Opens the Measured value window # dialog window, in which the
parameters for the selected window can be edited (see Chapter
5.6.3.2.4.9, page 536)
.

[Delete]

Deletes the window selected in the table.

Evaluation with volume window (mL)

[New]

Opens the Volume window # dialog window, in which the parameters
for a new window can be entered (see Chapter 5.6.3.2.4.10, page
537)
.

[Properties]

Opens the Volume window # dialog window, in which the parameters
for the selected window can be edited (see Chapter 5.6.3.2.4.10, page
537)
.

[Delete]

Deletes the window selected in the table.