3 det u, 1 det u - overview, Det u – Metrohm tiamo 2.4 Manual User Manual

Page 501

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

485

Fixed value

Value of the fixed endpoint.

Measured value
Input range

-20.000 - 20.000 pH

Time
Input range

0.0 - 999999.9 s

Volume
Input range

0.00000 - 9999.99 mL

5.6.3.2.3

DET U

5.6.3.2.3.1

DET U - Overview

Dialog window: Method

DET U Properties... DET U - 'Command name'

Command for Dynamic equivalence point titrations with potentio-
metric voltage measurement.

Devices
This command can be executed with the following instruments:

Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907

Titrino: 716, 721, 736, 751, 785, 794, 798, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Parameters
The parameters for the DET U command are set on the following seven
tabs:

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

Titration parameters
Parameters for the run of the titration.