3 det u, 1 det u - overview, Det u – Metrohm tiamo 2.4 Manual User Manual
Page 501
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
485
Fixed value
Value of the fixed endpoint.
Measured value
Input range
-20.000 - 20.000 pH
Time
Input range
0.0 - 999999.9 s
Volume
Input range
0.00000 - 9999.99 mL
5.6.3.2.3
DET U
5.6.3.2.3.1
DET U - Overview
Dialog window: Method
▶ DET U ▶ Properties... ▶ DET U - 'Command name'
Command for Dynamic equivalence point titrations with potentio-
metric voltage measurement.
Devices
This command can be executed with the following instruments:
Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907
Titrino: 716, 721, 736, 751, 785, 794, 798, 799
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the DET U command are set on the following seven
tabs:
■
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
■
Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.
■
Titration parameters
Parameters for the run of the titration.