tiamo determines the calibration data for the concentration measurement
as for the pH calibration in accordance with the number of the standard
solutions used.

1-point calibration

s: the existing value is retained, if available. If not, the theoretical

slope U

N

/ z is used.


c

B

: is set equal to zero.

Two-point calibration


c

B

: is set equal to zero.

Three- and multiple-point calibration
With at least three calibration solutions the calibration function is
determined according to the method of least squares. The sought cali-
bration parameters s, E(0) and c

B

are calculated iteratively. If the value

for c

B

is lower than approx. 0.2% of the lowest concentration, it is set

equal to zero. In such a case a linear relation between U

i

and log(c

i

)

results.
With four or more calibration standards, the variance Var is always cal-
culated and output. If c

B

= 0, the variance is even calculated for three

standards.

U

i

: measured potential

U

i,c

: potential calculated from the calibration curve

n: number of measuring points
F: degrees of freedom (2 for c

B

= 0, 3 for c

B

≠ 0)

Command variables
The following command variables are generated by the CAL LOOP Conc
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':

Identifi-
cation

Description

.BLV

Blank value calculated from the calibration

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
never been started