5 Method

tiamo 2.4

■■■■■■■■

479

Break-point evaluation

on | off (Default value: off)

A break-point evaluation is used to determine sharp changes of direction
in the titration curve.

EP criterion

Measure of the minimum sharpness of the break-point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value
related to the total measured value alteration, even small changes in the
measured value can be evaluated as a break-point for a small measured
value range.

Input range

0 - 1.0

Default value

0.3

Slope

Minimum difference between the slope before and after the break-point.
The smaller the difference, the more break-points will be found.

Input range

0.0 - 10.0

Default value

0.9

Smoothing factor

The higher the smoothing factor, the fewer break-points will be found.

Input range

2 - 20

Default value

5

Window

A range (window) can be defined on the measured value axis, on the vol-
ume axis or on the time axis. The break-point evaluation will only be car-
ried out in the defined window. Only the first break-point in the defined
window will be recognized.

Selection

Measured value | Volume | Time | off

Default value

off

Lower limit

Value for the lower limit of the window.

Window = Measured value
Input range

-20.000 - 20.000 pH

Default value

-20.000 pH

Window = Volume
Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL