Metrohm tiamo 2.4 Manual User Manual

Page 898

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

882

■■■■■■■■

tiamo 2.4

Titrando, 855
Input range

–20.000 - 20.000

Default value

20.000

Titrino
Input range

–20.00 - 20.00

Default value

20.00

Upper hysteresis pH

Upper hysteresis of the measured value. If the measured value once again
falls below the upper limit by this hysteresis value, then the event Mea-
sured value upper limit OK
is triggered.

Titrando, 855
Input range

0.000 - 20.000

Default value

0.020

Action

Selection of the event to be triggered if the upper or lower limit is viola-
ted:

Selection

Cancel determination | Cancel command
(Titrando only)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT pH command will be canceled, then the exit track
(if present) will be started and the determination will be finished.
Cancel command (Titrando only)
The running STAT pH command will be canceled, then the next com-
mand will be carried out.
Wait for [Continue]
Reagent dosing in the running STAT pH command will be interrupted
and a message will be displayed. As soon as the monitored measured
value is again within the limits (including hysteresis), reagent dosing
can be resumed by pressing [Continue] in this message box.
Wait for limit ok
Reagent dosing in the running STAT pH command will be interrupted.
As soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are violated.

Advertising