3 meas opt - measuring parameters – Metrohm tiamo 2.4 Manual User Manual

Page 1065

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

1049

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.

Stirrer
The Stirrer section is only displayed for Device type = 089 Photometer.

Stirrer

Selection of the stirrer.

Selection

1 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate.

Input range

0 - 15

Default value

5

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.

5.6.4.14.3

MEAS Opt - Measuring parameters

Tab: Method

MEAS Opt Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measuring parameters for the Avantes spectrometer

Wavelength

Wavelength at which the measured quantity (absorbance) is recorded.

Input range

100.0 - 2,000.0 nm

Default value

400.0 nm