5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1062

■■■■■■■■

tiamo 2.4

5.6.4.15.3

MEAS Opt Conc - Measuring parameters

Tab: Method

MEAS Opt Conc Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

Measuring parameters for the Avantes spectrometer
The parameters for the measuring procedure can be defined on this tab.

Wavelength

Wavelength at which the measured quantity (absorbance) is recorded.

Input range

100.0 - 2,000.0 nm

Default value

400.0 nm

Blank value

Value by which the measured quantity (concentration) is corrected. Here
an expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured concentration
value can then be corrected.

Input range

–1E+12 - 1E+12 (max. 10 digits)

Default value

1.000

Sample size unit

Concentration unit for the blank value.

Entry

1 - 10 characters

Selection

mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | µg/mL | ppm | % | mEq/L

Default value

ppm

Integration time

Integration time for absorbance measurement.

Input range

0.01 - 600,000 ms

Default value

6 ms

Averaged spectra

Number of spectra that are recorded and averaged.

Input range

1 - 10000

Default value

10