3 kfc - start conditions, 4 kfc - control parameters, Start conditions – Metrohm tiamo 2.4 Manual User Manual

Page 850: Control parameters

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

834

■■■■■■■■

tiamo 2.4

851, 852
Input range

-15 - 15

Default value

8

Switch on/off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command.

5.6.3.6.3

KFC - Start conditions

Tab: Method

KFC Properties... Start conditions

Command name

Name of the command.

Entry

25 characters

Pause

Pause

Waiting time before the start of the titration.

Input range

0 - 999999 s

Default value

0 s

5.6.3.6.4

KFC - Control parameters

Tab: Method

KFC Properties... Control parameters

Command name

Name of the command.

Entry

25 characters

Endpoint

EP with

Measured value for the endpoint.

756, 831
Input range

-2000.0 - 2000.0 mV

Default value

50.0 mV

851, 852
Input range

-1250.0 - 1250.0 mV

Default value

50.0 mV