5 Method

tiamo 2.4

■■■■■■■■

921

not defined
If the sensor is not defined, then it will be checked at the start of the
determination whether a sensor of the type Thermoprobe is con-
nected to the measuring input. If this is the case, the connected sensor
is used; if it is no the case, an error message appears.

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Input range

-15 - 15

Default value

8

Switch on/off automatically

on | off (Default value: on)

If this option is activated, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command.

5.6.3.9.3

TET - Start conditions

Tab: Method

TET Properties... Start conditions

Command name

Name of the command.

Entry

25 characters

The start conditions are processed in the listed sequence before the titra-
tion is started.

Start volume

Start volume

Volume to be dosed with the indicated dosing rate before the start of the
titration.

Input range

0.00000 - 50.0000 mL

Default value

0 mL