Metrohm tiamo 2.4 Manual User Manual

Page 1087

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

1071

Input range

0 - 999,999 s

Default value

154 s

Measuring interval

Time interval for entering a measured value in the measuring point list.

Input range

0.25 - 999,999.00 s (Increment: 0.25)

Default value

2.00 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Input range

0 - 4,000 mAU

Selection

off

Default value

off

Temperature

Temperature

Measuring temperature, which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set to auto-
matic
or continuous on the General/Hardware tab under Sensor,
then the temperature will be measured continuously.

Input range

20.0 - 60 °C

Default value

25.0 °C

Measurement without drift control

Measuring time

Duration of the absorbance measurement.

Input range

0 - 999,999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Input range

0.25 - 999,999.00 s (Increment: 0.25)

Default value

2.00 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Input range

0 - 4,000 mAU

Selection

off

Default value

off