5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

678

■■■■■■■■

tiamo 2.4

Device type

Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.

If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.

Only editable for Device name = not defined, otherwise inactive.

Selection

856 Conductivity Module

Default value

856 Conductivity Module

Dosing device

Dosing device

Selection of the number of the dosing device (exchange or dosing unit)
with which the solution is to be dosed. All the dosing device connectors
which are possible with the selected device type are always displayed.

Selection

1 | 2 | 3 | 4

Default value

1

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Entry

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Selection

1

Default value

1