5 Method

tiamo 2.4

■■■■■■■■

695

SET Upol
Amperometric measurement with selectable polarization voltage (mea-
sured quantity current I).

SET Ipol
Voltametric measurement with selectable polarization current (mea-
sured quantity voltage U).

5.6.3.4.2

SET - Control range

Menu item: Method

SET Properties... Control parameters

Set a large control range for steep curves and a small control range for flat
ones. A good approximation for the start of the control range is given by
the point where the tangents intersect.

5.6.3.4.3

SET pH

5.6.3.4.3.1

SET pH - Overview

Dialog window: Method

SET pH Properties... SET pH - 'Command name'

Command for endpoint titrations with potentiometric pH measure-
ment.

Devices
This command can be executed with the following instruments:

Titrando: 808, 809, 835, 836, 841, 842, 857, 888, 901, 902, 904, 905,
906, 907

Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798,
799

Robotic Titrosampler: 855

Appearance
The command has the following appearance: