2 elt loop - properties – Metrohm tiamo 2.4 Manual User Manual
Page 1139
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1123
Identi-
fica-
tion
Description
.ETR
Result of the electrode test:
0 = No data available
1 = Short circuit
2 = Excessive start drift
3 = Wrong buffer
4 = Problems with the diaphragm
5 = Reference system defective
6 = No buffer values for the temperature or excessive devia-
tion of the temperature between buffers
7 = Partial short circuit
8 = Glass defective
9 = Poor electrode
10 = Usable electrode
11 = Good electrode
12 = Excellent electrode
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command has
never ended
.LCO
Loop counter = current number of completed runs, both of
Repeat loops and of While loops
.LST
Start time of the loop command (Date/Time)
.SLO
Electrode slope of the sensor used for the command (in mV
for ISE sensors)
.VAR
Variance of the calculated result of the calibration in the
selected unit
5.6.5.8.2
ELT LOOP - Properties
Dialog window: Method
▶ ELT LOOP ▶ Properties...
Command name
Name of the command.
Entry
25 characters