8 Manual control
tiamo 2.4
■■■■■■■■
1621
Input range
100.0 - 2000.0 nm
Default value
1000.0 nm
Integration time
Integration time for spectrum recording.
Input range
0.01 - 600000 ms
Default value
6 ms
Averaged spectra
Number of spectra that are recorded and averaged.
Input range
1 - 10000
Default value
10
Smoothing
Number of neighboring pixels with which the value for each measurement
pixel is determined.
Input range
0 - 100 Pixels
Default value
0 Pixels
Flash rate
Number of flashes per second of a pulsed xenon lamp as light source.
Input range
0 - 100 Hz
Selection
off
Default value
off
off
For the continuous type of the light source.
[Start]
Starts measurement of the intensity.
[Stop]
Stops measurement of the intensity.
8.4.5.4
Measure - Sample spectrum
Tab: Manual
▶ Sample spectrum
Here a sample spectrum can be measured and displayed.