5 Method
tiamo 2.4
■■■■■■■■
757
Input range
0.0 - 99.9 µL/min
Default value
0.0 µL/min
Stop volume
Stops when the given volume has been added after the start of the titra-
tion. The stop volume should be adapted to the size of the titration cell in
order to prevent any overflow.
Titrando, 855
Input range
0.000000 - 9999.99 mL
Default value
20.00 mL
Selection
off
Default value
off
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Titrando, 855
Input range
1 - 999999 s
Selection
off
Default value
off
Stabilizing time
Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.
Input range
0 - 999 s
Default value
0 s
Automatic start after sample addition time
on | off (Default value: on)
If this option is activated, then the titration will be started automatically
after the sample addition time.
Sample addition time
Waiting time between stop of conditioning and start of titration during
which the sample can be added.
Input range
0 - 99 s
Default value
0 s