5 Method

tiamo 2.4

■■■■■■■■

757

Input range

0.0 - 99.9 µL/min

Default value

0.0 µL/min

Stop volume

Stops when the given volume has been added after the start of the titra-
tion. The stop volume should be adapted to the size of the titration cell in
order to prevent any overflow.

Titrando, 855
Input range

0.000000 - 9999.99 mL

Default value

20.00 mL

Selection

off

Default value

off

Stop time

Determination stops when the specified time has elapsed since the start of
the conditioning.

Titrando, 855
Input range

1 - 999999 s

Selection

off

Default value

off

Stabilizing time

Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.

Input range

0 - 999 s

Default value

0 s

Automatic start after sample addition time

on | off (Default value: on)

If this option is activated, then the titration will be started automatically
after the sample addition time.

Sample addition time

Waiting time between stop of conditioning and start of titration during
which the sample can be added.

Input range

0 - 99 s

Default value

0 s