4 send, 1 send - overview, Send – Metrohm tiamo 2.4 Manual User Manual
Page 1281
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1265
*******1 expects an active input line 0 (1 = set). For example, this line is
set by a Titrino after a titration has been completed and the Titrino can
accept a start signal again.
NOTE
Input lines that are of no interest or for which no defined condition can
be predicted should also be masked with an asterisk *.
Max. waiting time
on | off (Default value: off)
If this option is activated, then a maximum waiting time can be entered
for the input signal.
Input range
0.0 - 9999.9
Default value
10.0
Selection
min | s
Default value
min
NOTE
If the option Max. waiting time is not activated, it is being waited
endlessly for the input signal.
5.6.9.4
SEND
5.6.9.4.1
SEND - Overview
Dialog window: Method
▶ SEND ▶ Properties... ▶ SEND - 'Command name'
Command for Sending event messages to commands, tracks or to
tiamo.
Appearance
The command has the following appearance:
Parameters
The parameters for the command SEND are configured in the following
dialog window:
■