5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1064

■■■■■■■■

tiamo 2.4

Measurement with drift control

Signal drift

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Input range

0.1 - 999,000 mAU/min

Default value

1.0 mAU/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the minimum waiting time is elapsing.

Input range

0 - 999,999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered, then a waiting time that is suitable for
the drift will be calculated automatically according to the following for-
mula:

Input range

0 - 999,999 s

Default value

154 s

Measuring interval

Time interval for entering a measured value in the measuring point list.

Input range

0.1 - 999,999.0 s

Default value

2.0 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Input range

0 - 4,000 mAU

Selection

off

off
No stop.

Measurement without drift control

Measuring time

Duration of the absorbance measurement.