5 Method

tiamo 2.4

■■■■■■■■

885

Monitoring Temperature

Monitoring Temperature

on | off (Default value: off)

If this option is enabled, then the temperature will be monitored and any
violated limits will be entered in the measuring point list.

Lower limit

Lower limit of the temperature. If the temperature falls below this limit,
the event Lower temperature limit violated is triggered.

Titrando, 855
Input range

–20.0 - 150.0 °C

Default value

–20.0 °C

Titrino
Input range

–170.0 - 500.0 °C

Default value

–170.0 °C

Lower hysteresis

Lower hysteresis of the temperature. If the temperature exceeds the lower
limit by this hysteresis value again, then the event Lower temperature
limit OK
is triggered.

Titrando, 855
Input range

0.0 - 150.0 °C

Default value

0.2 °C

Upper limit

Upper limit of the temperature. If the temperature exceeds this limit, then
the event Upper temperature limit violated is triggered.

Titrando, 855
Input range

–20.0 - 150.0 °C

Default value

150.0 °C

Titrino
Input range

–170.0 - 500.0 °C

Default value

500.0 °C

Upper hysteresis

Upper hysteresis of the temperature. If the temperature once again falls
below the upper limit by this hysteresis value, then the event Upper tem-
perature limit OK
is triggered.