3 meas tmf - measuring parameters – Metrohm tiamo 2.4 Manual User Manual

Page 1085

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

1069

Selection

089 Photometer

Default value

089 Photometer

Sensor

Temperature measurement

Type of temperature measurement.

Selection

continuous | off

continuous
The temperature will be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate.

Input range

0 - 15

Default value

5

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.

5.6.4.16.3

MEAS TMF - Measuring parameters

Tab: Method

MEAS TMF Properties… Measuring parameters

Command name

Name of the command.

Entry

25 characters