3 meas tc cond - measuring parameters – Metrohm tiamo 2.4 Manual User Manual

Page 1051

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

1035

Switch off automatically

on | off (Default value: on)

If this option is activated, the stirrer will be switched off automatically
when the command has finished.

5.6.4.11.3

MEAS TC Cond - Measuring parameters

Tab: Method

MEAS TC Cond Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement frequency

712
Selection

auto | 300 Hz | 2.4 kHz

Default value

auto

auto
The more suitable of the two frequencies is selected.

Stop time

Maximum period of time for measurement.

Input range

0 - 999999 s

Default value

3000 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

712
Input range

0.4 - 999999 s

Default value

2.0 s

856
Input range

0.1 - 999999 s

Default value

2.0 s

Sample solution

Selection of the sample solution from the table Sample solutions (TC
conductivity)
.

Selection Table entries
Default value first entry of table