5 Method
tiamo 2.4
■■■■■■■■
1035
Switch off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched off automatically
when the command has finished.
5.6.4.11.3
MEAS TC Cond - Measuring parameters
Tab: Method
▶ MEAS TC Cond ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement frequency
712
Selection
auto | 300 Hz | 2.4 kHz
Default value
auto
auto
The more suitable of the two frequencies is selected.
Stop time
Maximum period of time for measurement.
Input range
0 - 999999 s
Default value
3000 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
712
Input range
0.4 - 999999 s
Default value
2.0 s
856
Input range
0.1 - 999999 s
Default value
2.0 s
Sample solution
Selection of the sample solution from the table Sample solutions (TC
conductivity).
Selection Table entries
Default value first entry of table