5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

832

■■■■■■■■

tiamo 2.4

Selection

Sensor name | Metal electrode

Default value

Metal electrode

I(pol)

Polarization current at the indicator electrode.

Selection

2 µA | 5 µA | 10 µA | 20 µA | 30 µA

Default value

10 µA

NOTE

When changing this parameter, make sure that the polarization current
of the indicator electrode, the endpoint and the control range are
linked closely with one another (see Chapter 5.6.3.6.4, page 834).

Electrode check

on | off (Default value: on)

If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement.

851, 852
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will be used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.

Cell

Generator type

Selection of the generator type.