5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

896

■■■■■■■■

tiamo 2.4

not defined
No tests will be carried out.

Tandem dosing

on | off (Default value: off)

If this option is enabled, the dosing is carried out without interruption
with a combination of two dosing devices so that the second dosing
device is dosing while the first one is being filled and vice-versa (see Chap-
ter 5.6.3.8.5, page 863)
. This parameter is displayed only for Titrandos
and 855.

Dosing device

Selection of the number of the dosing device (exchange or dosing unit)
with which the solution is to be dosed. All the dosing device connectors
which are possible with the selected device type are always displayed.

Titrando
Selection

1 | 2 | 3 | 4

Default value

2

855
Selection

1 | 2 | 3

Default value

2

Solution

Entry of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method run to see whether the
correct solution has been attached to the selected dosing device and
whether the dosing device type is correct. With nonintelligent exchange
or dosing units, only the cylinder volume is checked.

Titrando, 855
Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Filling rate

Rate at which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing unit used. If the entered filling rate is too high for
the selected dosing device, it will automatically be reduced during filling
to the largest possible value (see Chapter 5.6.3.8.5, page 863).