7.9 774 Oven Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

1534

■■■■■■■■

tiamo 2.4

Max. temperature

With the 774 Oven Sample Processor display of the maximum tempera-
ture set on the device only.

With the 774 Oven Sample Processor display only.
Input range

50 - 250 °C

Default value

250 °C

Temperature correction

Enter the temperature correction. This parameter directly affects the tem-
perature control and allows it to be adjusted. In this way any temperature
difference that may occur between the oven and sample temperature can
be compensated.

With the 774 Oven Sample Processor display of the temperature correc-
tion set on the device only.
Input range

-10 - +10 °C

Default value

0 °C

7.9.7

Properties - Gas

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Gas

On the tab Gas the parameters for the gas flow on the 774 Oven Sam-
ple Processor
are shown. The parameters cannot be edited, they are
only read off by the device.

Unit for gas flow

Shows the unit selected at the device for the gas flow display (mL/min or
L/h).

Minimum gas flow

Shows the lower warning limit set on the device for the gas flow.

Maximum gas flow

Shows the upper warning limit set on the device for the gas flow.

NOTE

If the minimum gas flow is undercut or the maximum gas flow is
exceeded then an error message will appear.