5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

818

■■■■■■■■

tiamo 2.4

Stop criterion

Stop criterion

The titration is canceled after the last dosing once the endpoint has been
reached and the stop criterion has been fulfilled. The titration can be can-
celed when a particular Drift has been reached or after a preset Time.

Titrino
Selection

Drift | Time

Default value

Drift

Titrando
Selection

Drift | rel. drift | off | Time

Default value

Drift

Drift
The titration is canceled when the Stop drift has been reached.
rel. drift
The titration is canceled when the relative stop drift (drift value at the
start + Relative stop drift) has been reached.
off
The titration is not canceled.
Time
The endpoint must be exceeded during a certain period of time: the
delay time. The same delay time with different smallest possible vol-
ume increments (depending on the volume of the exchange or dosing
unit) equals different switch-off points.

NOTE

The stop conditions always lead to a stop, even if the stop criterion has
not been reached or has been switched off.

Stop drift

Titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.

Input range

1 - 999 µL/min

Default value

20 µL/min

Relative stop drift

The titration is canceled when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift value at the titration start
and the relative stop drift defined here. This parameter is displayed only if
Stop criterion = rel. drift.