5 meas t - additional measured values, Additional measured values – Metrohm tiamo 2.4 Manual User Manual

Page 1018

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1002

■■■■■■■■

tiamo 2.4

Lower limit

Value for the lower limit of the window.

Time for the lower limit of the window.

Window = Measured value
Input range

-20.0 - 150.0 °C

Default value

-20.0 °C

Window = Time
Input range

0 - 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Value for the upper limit of the window.

Window = Measured value
Input range

-20.0 - 150.0 °C

Default value

150.0 °C

Window = Time
Input range

1 - 999999 s

Default value

999999 s

5.6.4.7.5

MEAS T - Additional measured values

Tab: Method

MEAS T Properties... Additional measured values

Command name

Name of the command.

Entry

25 characters

A maximum of 6 additional measured values can be defined here which
can then be saved together with the measured values present in the
default settings in additional measured value columns Calculated # or
External # in the measuring point list.

Additional calculated measured values

Additional calculated measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be
presented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.