10 rls dev, 1 rls dev - overview, Rls dev – Metrohm tiamo 2.4 Manual User Manual
Page 1255
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1239
ted in the manual. For the 874 USB Oven Sample Processor, these values
can be found in the table Gas flow factors of various gases.
Input range
0.001 - 2.000 (Increment: 0.001)
Default value
1.000
Gas flow
Inlet
Selection of the carrier gas inlet.
Selection
Pump | Valve
Default value
Pump
Pump
Use of ambient air as carrier gas.
Valve
Use of gas out of a compressed gas cylinder.
Flow rate
874
Input range
10 - 150 mL/min
Default value
50 mL/min
Gas flow
Selection
Switch on | Switch off
Default value
Switch on
Switch on
Switches the gas flow on.
Switch off
Switches the gas flow off.
5.6.7.10
RLS DEV
5.6.7.10.1
RLS DEV - Overview
Dialog window: Method
▶ RLS DEV ▶ Properties... ▶ RLS DEV - 'Device name'
Command to release a device for using it in other parallel running meth-
ods. It removes the reservation made by the current method.
Devices
This command can be executed with the following devices:
Sample Processor: 730, 774, 778, 789, 814, 815, 864, 874
Robotic Titrosampler: 855