5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1178

■■■■■■■■

tiamo 2.4

time * dosing rate and is not shown (see Chapter 5.6.6.3.2, page
1166)
.

Selection

Volume/dosing rate | Volume/dosing time |
Dosing rate/dosing time

Default value

Volume/dosing rate

Volume

Fixed volume that is to be dosed. This parameter is only displayed for
Dosing criterion = Volume/Dosing rate or Volume/Dosing time.

Input range

0.00000 - 99999.9 mL

Default value

10.0000 mL

Dosing rate

Rate at which dosing should take place. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing device
used. If the entered dosing rate is too high for the selected dosing device,
it will automatically be reduced during dosing to the largest possible
value. This parameter is only displayed for Dosing criterion = Volume/
Dosing rate
or Dosing rate/Dosing time.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Default value

maximum mL/min

Titrino
Input range

0.01 - 150.00 mL/min

Default value

maximum mL/min

NOTE

The dosing rate should be reduced for viscous liquids.

Dosing time

Time during which dosing is to be carried out. This parameter is only dis-
played for Dosing criterion = Volume/Dosing time or Dosing rate/
Dosing time.

Input range

0 - 9999999 s

Default value

100 s

Pause

Waiting period before dosing is started.