5 Method

tiamo 2.4

■■■■■■■■

851

time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.

Input range

0 - 999999 s

Default value

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed.

Input range

0 - 999999 s

Default value

15 s

Pause

Pause

Waiting time before the start of the titration.

Input range

0 - 999999 s

Default value

0 s

5.6.3.7.4

BRC - Control parameters

Tab: Method

BRC Properties... Control parameters

Command name

Name of the command.

Entry

25 characters

Endpoint

EP with

Measured value for the endpoint.

Input range

-1250.0 - 1250.0 mV

Default value

200.0 mV

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.

Selection

slow | optimal | fast | User

Default value

optimal