7 brc - conditioning, Conditioning – Metrohm tiamo 2.4 Manual User Manual

Page 871

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

855

5.6.3.7.7

BRC - Conditioning

Tab: Method

BRC Properties... Conditioning

Command name

Name of the command.

Entry

25 characters

Conditioning

on | off (Default value: on)

Enable/Disable the conditioning. If the conditioning is disabled, the follow-
ing parameters are not visible.

Start drift

The determination can only be started when the currently measured drift
is less than the start drift entered here. In such cases, the Status in the
Run subwindow displayed on the Single determination or Determina-
tion series
tab will then switch to COND READY.

Input range

1 - 999 µg/min

Default value

20 µg/min

Drift correction

If this parameter is set to automatic or manual the drift correction ( =
Drift value × Time for drift correction) caused by empty consumption
during titration is deducted from the total consumption. The Time for
drift correction
is the period between end of conditioning and end of
titration.

Selection

automatic | manual | off

Default value

off

automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.

Drift value

Drift value for drift correction. This parameter is only shown for Drift cor-
rection
= manual.

Input range

0.0 - 999.9 µg/min

Default value

0.0 µg/min

Advertising