Metrohm tiamo 2.4 Manual User Manual

Page 1110

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1094

■■■■■■■■

tiamo 2.4

Max. waiting time

If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered then a waiting time that is suitable for
the drift will be calculated automatically according to the following equa-
tion:

Input range

0 - 999999 s

Default value

52 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Input range

0.1 - 999999.0 s

Default value

2.0 s

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.

Input range

-20.0 - 150°C

Default value

25.0

Switch off stirrer during measurement

Switch off stirrer during measurement

on | off (Default value: off)

If this option is activated, then the stirrer is switched off during the mea-
surement. In this case the following two parameters can be edited:

Stir before measurement

Stirring time between end of dosing and measurement.

Input range

0 - 999999 s

Default value

0 s

Pause before measurement

Waiting time before the measurement during which no stirring takes
place.

Input range

0 - 999999 s

Default value

0 s

Advertising