Metrohm tiamo 2.4 Manual User Manual
Page 962
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
946
■■■■■■■■
tiamo 2.4
Titrando, 855, 867
Input range
0.1 - 999999.0 s (Increment: 0.1)
Default value
2.0 s
Titrino
Input range
0.08 - 16200 s (Increment: 0.08)
Default value
2.0 s
Stop measured value pH
Stops when the preset measured value has been reached since the start of
the measurement.
Titrando, 855, 867
Input range
-20.000 - 20.000
Selection
off
Default value
off
off
No stop.
Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:
Measuring time
Maximum period of time for measurement.
Titrando, 855, 867
Input range
0 - 999999 s
Default value
120 s
Titrino
Input range
0 - 9999 s
Default value
120 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Titrando, 855, 867
Input range
0.1 - 999999.0 s (Increment: 0.1)
Default value
2.0 s
Titrino
Input range
0.08 - 16200 s (Increment: 0.08)
Default value
2.0 s