5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

946

■■■■■■■■

tiamo 2.4

Titrando, 855, 867
Input range

0.1 - 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Input range

0.08 - 16200 s (Increment: 0.08)

Default value

2.0 s

Stop measured value pH

Stops when the preset measured value has been reached since the start of
the measurement.

Titrando, 855, 867
Input range

-20.000 - 20.000

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:

Measuring time

Maximum period of time for measurement.

Titrando, 855, 867
Input range

0 - 999999 s

Default value

120 s

Titrino
Input range

0 - 9999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Titrando, 855, 867
Input range

0.1 - 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Input range

0.08 - 16200 s (Increment: 0.08)

Default value

2.0 s