5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1086

■■■■■■■■

tiamo 2.4

the drift will be calculated automatically according to the following equa-
tion:

Input range

0 - 999999 s

Default value

52 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Input range

0.1 - 999999.0 s

Default value

2.0 s

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.

Input range

-20.0 - 150°C

Default value

25.0

Switch off stirrer during measurement

Switch off stirrer during measurement

on | off (Default value: off)

If this option is activated, then the stirrer is switched off during the mea-
surement. In this case the following two parameters can be edited:

Stir before measurement

Stirring time between end of dosing and measurement.

Input range

0 - 999999 s

Default value

0 s

Pause before measurement

Waiting time before the measurement during which no stirring takes
place.

Input range

0 - 999999 s

Default value

0 s