5 Method

tiamo 2.4

■■■■■■■■

1165

Titrando, 814, 815, 855, 864, 874
Input range

0.01 - 166.00 mL/min

Default value

maximum mL/min

730, 774, 778, 789
Input range

0.01 - 160.00 mL/min

Default value

maximum mL/min

Titrino
Input range

0.01 - 150.00 mL/min

Default value

maximum mL/min

NOTE

The filling rate should be reduced for viscous liquids.

Fill automatically at start

on | off (Default value: on)

If this option is activated, the buret is filled automatically before dosing.
This parameter is displayed only for Titrandos, 846, 855, 856, 864, 867,
814, 815 and 874.

This option ensures that the buret is filled and set to the dosing port of
the selected solution before the ADD command is carried out.

NOTE

If Fill automatically at start is disabled, then the ADD command
starts adding solvent immediately. It has to be ensured in the method
(e.g. by adding a preceding ADD or LQH command) that the buret is
set to the correct dosing port.

Fill automatically at end

on | off (Default value: on)

If this option is activated, the buret is filled automatically after dosing.

This option ensures that the buret is filled again after the ADD command
has been carried out and set to port 2, so that it can be removed.